Compliance Standards:
IS 1209, IS 1448: Part 21, IS 1448: Part 66, ASTM D93, ISO 2719, IP 34, JIS K2265-3
Salient Features:
- Standalone model with touch screen display.
- Accurate determination of Flash Point
- Easy operation: Select procedure, enter Expected Flash Point and click on start. Instrument follows the exact procedure and completion of the test cycle is signaled by beeps.
- Rapid heating procedure for high flash point till selected temperature below expected flash point
- Search programs enable a quick determination of fully unknown samples
- Interchangeable gas or electric igniter source
- Automatic barometric pressure correction Interlocks for safe operation
User Interface:
- Windows based user friendly Software
- In-built method and results storage
- Standalone with In Built PC & 8.4” touch screen display
- With LIMS compatibility RS 232, LAN , Wi-Fi, Bluetooth and USB Ports for Printer & other peripherals.
Heating:
- Band type heater, 450 watt @ 230VAC
- Heat rate as per method or defined by the user from 0.5 to 14°C/min
- Accuracy : 0.02 °C
- Resolution : 0.01 °C
Cooling : Centrifugal Blower 56.5 CFM, 35 watt @ 230VAC
Stirring : Stepper Motor controlled Speed as per method or user defined from 20 to 300 rpm
Ignition:
- DC motor controlled movement
- S/W selectable Electric/gas igniter & 0.5 to 5°C dipping interval
- Precise Gas flow and 3 step Electric Igniter brightness adjustment
Arm Movement
- Automatic DC motor controlled swing arm
- Permanently fixed Cup led with stirrer, sample temperature sensor, Flash detector and Igniter for easy operation
Measurements
- Auto-Cal : Prior to Every run, auto temperature calibration will be performed without the use of any calibration device.
- Sample Temperature : PT-100 Class A, SS sheath, Range 0 to 400 °C
- Bath Temperature : PT-100 Class A, Disk Type, Range 0 to 500 °C
- Flash Detection : Thermal detection by Thermocouple.
- Instrument Measurement range : Ambient to 400 °C
- Amb. Pressure : Built in pressure sensor, Range 700 to 1100 hpa (500 to 800 mmHg)
Safety interlocks:
- Temperature sensors (bath/sample) is found defective.
- Flash Detector is found defective.
- Electric Igniter is blown off.
- Test Stop at above expected FP as selected or at 400 °C
- Upon Detection of Fire by thermal sensor, warning buzzer will beep and power will cut off
- Test cover is not set in place.
Operating requirements:
- Power Supply : 230 Volts, 50 Hz, 1000 watt
- Housing : (W410 x D420 x H330) mm
- Weight : 16 Kg
- Ambient Temp : 10 to 35 °C
- Relative Humidity : 80% at 35 °C