Construction Details:
- CE Certified
- Designed as per ICH guidelines to meet, WHO and USFDA requirement
- Fulfills storage conditions of 25° C - 60% RH, 40° C - 75% RH, 30° C - 65% RH, 25° C - 40% RH, 30° C - 35% RH
- Chamber equipped with PLC & Touch screen HMI system
- Highly accurate Individual control Loop for precise control of temp and humidity
- Interior fabricated from mirror polished Stainless Steel (S.S. - 304)
- Exterior body is fabricated from Stainless Steel (S.S. - 304)
- Toughened glass wondow to inspect samples without affecting the chamber temperature.
- Prefabricated Puf Panels with cam-locks to avoid thermal leakage
- Electrical wiring as per CE Compliances
- Chamber illumination by LED light with door switch.
- User friendly and tactfully designed chamber door and locking mechanism
- Facility to open chamber door from inside, in case of emergency
- Aesthetic outer appearance and high quality
- Supplied complete with IQ/OQ/PQ Documents, Calibration certificates and Instruction manual.
Following advance features are available:
- Equipped with 7" Touch screen HMI Controller and PLC System
- Equipped with GSM MODULE (Mobile alert configuration) system for parameter deviation
- Equipped with Magnetic Door Lock Mechanism
- Password protected door access system
- Supplied complete with 21-CFR (PART-11) Compliance PC Software
- Audit trail generation in PC Software
- Equipped with In-Built Data logger (Uniformity module) system
- Also equipped with 4 Temp + 4 RH Sensor for data logging purpose
- Monitor Low water level status
- Facilitates automatic water filling system
- Auto resume after power failure
- Ready to use STANDBY REFRIGERATION SYSTEM in case main system fails
- Ready to use STANDBY HUMIDITY SYSTEM in case main system fails
- Auto Switch-Over facility from Mains to Standby system and vice versa
- Calendar options for choosing the days on which Either Mains or Standby system will remain active
Salient Features of 21-CFR Compliance Software:
- Multi Chamber data - Acquisition Software
- Generates graphical analysis of DATA as per - 21 CFR Compliance guidelines
- File encryption Facility
- Secure login facility
- Password protected. System access is controlled by the system supervisor
- Limit system access
- Limit module access
- Password expiration
- Bilateral user ID and password identification
- Access failure lock out
- System access logs
- System audit trail report.
- System Input / Output logs
- Mean kinetic temperature calculation
- Maintains log of user activities like users logged in, login/logout time, and changes made by the user in software settings Audit Trail Reports.
- Ensure Gap-free Records
- Schedule Reports via Email
- Meet Tough Standards
- View Environments in Real-Time
- Flexible Alarm Management
- Consistent Alarm Acknowledgements
- Highly Secure & Temper proof
- Protection From Data Loss
Air Circulation:
Temperature is maintained by a quiet running blower circulation throughout the chamber. Forced air circulated vertically down and re-circulated throughout the chamber for uniform temperature and humidity distribution.
Humidity Generation:
Humidity created by steam injection method. The boiler tank is fitted in the back side of the chamber for better servicing. Electromagnetic switch is provided for controlling the wet heater from burning off, if water level is not adequate. Float valve is provided to automatically control & maintain water level in the boiler tank.
Heating & Cooling System:
Long lasting SS tubular heaters used as heating element. The stainless steel fins provided to ensure better heat transfer. Hermetically sealed compressor CFC free compressor (134 A gas) coupled with evaporation coil and condenser. All refrigeration components like expansion valve, driers, oil separators, pressure gauges and vibration eliminators provided. Reputed company make refrigeration system such as Emerson/Kirloskar Make
Door & Gasket:
Robust type hinges fitted on the right hand side of the chamber with front opening single door. Leak proof double lined silicon gasket will be provided. The inner surface of the door will be provided with pre-polished stainless steel and will be mounted such that conduction is avoided.
Trays for samples:
- Removable Stainless steel rod trays
- Strong enough to withstand heavy load of test specimens
- Such trays will also facilitate better air circulation.
- The distance between trays will be 15-20 centimeters.
- Number of tray supplied : 36 Numbers
Electrical controls:
Relays, Contactors, SSR, MCB, Mains Switch, Fan control switch, compressor ON/OFF switch with indicator, Dry heater switch with indicator for safe and ease of operation.
Electrical Circuitry Wires:
CE Certified : Inner circuitry wires with individual number tags for ease of servicing and maintenance
Safety Provisions:
- Over temperature thermostat & thermal fuses provided for Heating system.
- Low water level boiler cut off
- Electrical overload cut off
- Time delay for compressor switch ON
- Electrical circuit breaker
Viewing window
Multi pane window, toughened clear glasses sealed with silicone and separated by spacers to form a completely vacuumised window.
Wiper
Viewing window facilitates with wiper to clear the condensed water droplets on the glass.
Chamber Size |
2.50 x 3.00 x 2.50 (W x D x H) Meter |
Chamber volume |
18750 liters |
Inner chamber |
Stainless steel S.S. - 316 Mirror Polished with cam lock |
Exterior body |
Stainless steel S.S. - 304 Matt finished |
Insulation |
Puff insulation |
Insulation thickness |
75mm |
No. of trays |
36 Numbers |
Controlling System |
Highly advanced Touch screen HMI & PLC Control system |
Display |
Digital LCD Display (7" Screen Size) |
Temperature Range |
10° C to 70° C |
Temperature Accuracy |
± 0.5° C |
Temperature sensor |
Pt - 100 sensor |
No. of Temperature Sensors |
05 Nos (01 Main + 04 For data logging purpose) |
Humidity Range |
40% RH to 90% RH |
Humidity Accuracy |
± 3% RH |
Rh sensor |
Capacitance type RH Sensor |
No. of Rh Sensors |
05 Nos (01 Main + 04 For data logging purpose) |
Validation port hole |
150 mm diameter, port hole provided with removable sealed cap. |
Observation window |
To inspect samples without disturbing chamber operation |
Illumination |
LED Light |
Time & Date |
User adjustable Date & Time settings |
PC Software |
21-CFR (Part-11) compliance software as per USFDA Guidelines |
Communication |
RS - 232/RS - 485 communication port |
Magnetic door lock |
Password protected entry levels for Operator, Supervisor etc |
IQ, OQ, PQ |
IQ, OQ, DQ, Documents as per USFDA Guidelines, Validation Report and Instruction Manual. |
Blower |
For uniform temp & humidity distribution |
Power Consumption |
3000 Watts (3 KW) |
Power Supply |
230 Volts, 50 HZ, Single Phase, AC Supply |