Complies with following International Standards
ASTM D892, IP 146, ISO 6247
Construction Details
The complete assembly consists of following parts and accessories.
1) Foaming Bath Apparatus - (Glass jar with four Test Cylinders and separate Heating & Cooling arrangement)
- Suitable for I, II & III sequence, As per ASTM D892 Standard.
- Temperature range: For bath I = 24°C ± 0.5°C, For bath II = 93.5°C ±0.5°C
- Consists of constant temperature transparent glass heating jar bath
- Heating glass jar will be of Heat resistant material having height of 450 mm and Diameter of 300 mm
- The transparent heating test bath will be provided with top cover
- One Tank cover will be having two hole of Ø 125 mm, which allows 1000 ml cylinders to get through it
- The heating glass jar is transparent and clear enough to permit observation of the graduations on the cylinder.
- Tactfully designed Stainless steel clamps, incorporated within tank cover, to hold the test cylinders in place.
- Microprocessor PID control provides quick temperature stabilization without overshoot and the bath is protected by an over temperature control circuit.
- Dual LED display provides actual and set point temperature values in °C/°F format.
- One glass jar is fitted with cooling coil, to which cold water will be circulated from the external chiller to conduct foaming operation on 24 °C temperature and another glass jar will be fitted with heating elements to conduct foaming experiment on 93.5 °C Temperature
- Heating supplied by an armored stainless steel heater in the test bath, to conduct the test on 93.5 °C
- Automatic boost heating
- Automatic high temperature & low liquid level safety cut‑off.
- Supplied with stirrer to maintain uniform temperature distribution throughout the bath.
- Both the Bottom bases house MICROPROCESSOR BASED DIGITAL PID controllers with over‑temperature alarm and probe PT ‑ 100 temperature sensor along with air compressor pump for supplying adequate air through flow meter to diffuser ball.
2) Test cylinders
- No. of cylinders supplied : 04 Numbers
- Material of construction : Borosilicate transparent
- Capacity : 1000 ml
- Graduation : 1‑1000 ml
- Cylinders will be circular at the top
- Cylinders will be fitted with a rubber stopper, having one hole at the center for the air‑inlet tube and a second hole off‑center for an air‑outlet tube.
3) Air (Gas) diffuser Stone
- Imported Gas Diffuser stone
- No. of Stones Provided : 04 Number
- Diameter : 25.4‑mm (1 inch)
- Maximum pore diameter : Not greater than 80 Microns
- Material of construction : Made of fused crystalline alumina grain
- Gas diffusers are attached to air‑inlet tubes by a suitable means.
4) Rubber tubing
- Supplied with 3 meter long rubber tubing for connection to gas diffuser stone and Air supply pump
5) Flow meter (Rotameter)
- No. of flow meter : 02 numbers
- Make : Of standard reputed make with calibration certificate
- Aair flow rate : 94 ± 5 mL/min through the gas diffuser.
6) Air compressor (Pump)
- No. of pumps supplied : 02 independent (oil free) blowing pumps
- Connected to : Flow meter to maintain air flow rate at 94 ± 5 mL/min
Note
- Low temperature of 24 °C can be achieved by adding ICE into the water or by placing the whole equipment in to the Air conditioning room & maintained at the above temperature OR by externally circulating the chilled water. Copper tubing is provided to facilitate the water circulation (Chiller quoted separately).
- Client must arrange external chiller to conduct the test on 24 °C temperature or can buy extra from us.