Compliance Standards:
ASTM D2699-86, ASTM D2700-86, ASTM D4737-03, ASTM D613, EN ISO 5165.
Details of the instrument:
- Advanced high precision microprocessor based design
- Comprises of 8-Keys antiglare keyboard for ease of operation
- Instrument is Sturdy, Wear-Proof and offers excellent resistance to aggressive environment
- Supports automatic computer based calibration facility
- Store the data in instrument memory, including time and date
- Can be interfaced to PC Software through USB Interface facility
- Data can be transferred to PC Software in Microsoft excel format or Text Format
- Intended for rapid petroleum products quality monitoring in field and laboratory conditions
Instrument Capabilities
- Determining octane numbers for motor petrols. Complies with ASTM D2699-86, ASTM D2700-86.
- Determining cetane numbers for diesel. Complies with ASTM D4737-03, ASTM D613, EN ISO 5165.
- Determining solidification temperature and diesel fuel type.
- Content of pour-point depressants for diesel fuels.
- Content of kerosine in diesel fuel.
Delivery Package
- Electronic computing module 1 piece
- Sensor #1 1 piece
- Sample imitator 1 piece
- Mini-USB cable 1 piece
- Power elements (can be removed) 4 pieces
- User manual 1 piece
- Warranty certificate 1 piece
- Quick start 1 piece
- Tool canvas bag 1 piece
Petrol Octane Number |
|
Measured petrols octane numbers range |
40–125 ON |
Acceptable limit of octane number measurement basic error, max |
± 0.5 ON |
Limit of acceptable difference between parallel octane number measurements, max |
± 0.2 ON |
Petrol Oxidation Breakdown Time |
|
Petrol oxidation breakdown time measurement range |
50-2400 Minutes |
Acceptable basic error limit of petrol oxidation breakdown time |
10 Minutes |
Cetane Number |
|
Cetane numbers measurement range |
20–100 CN |
Acceptable basic error limit of cetane numbers, max |
±1.0 CN |
Acceptable difference limit between cetane numbers parallel measurements, max |
± 0.5 CN |
Diesel Fuel Pour Point |
|
Acceptable error limit when determining diesel-fuel pour points |
± 5 °C |
Kerosine content determination range in diesel fuels |
0-95 % |
Acceptable basic error limit when determining kerosene content in diesel fuels |
3 % |
Measurement Time |
1–5 s |
Insufficient power supply indication operation threshold |
5.4 V |
Instrument useful life |
6 |
Overall Dimensions |
|
Electronic module |
100 x 210 x 40 mm |
Sensor #1 and #2 |
60 х 100 mm |
Weight |
680 Grams |