Petrol Octane Number |
|
Measured petrols octane numbers range |
40–135 ON |
Acceptable limit of octane number measurement basic error, max |
± 0.5 ON |
Limit of acceptable difference between parallel octane number measurements, max |
± 0.2 ON |
Petrol Oxidation Breakdown Time |
|
Petrol oxidation breakdown time measurement range |
50-2400 minutes |
Acceptable basic error limit of petrol oxidation breakdown time |
10 Minutes |
Cetane Number |
|
Cetane numbers measurement range |
20–100 CN |
Acceptable basic error limit of cetane numbers, max |
±1.0 CN |
Acceptable difference limit between cetane numbers parallel measurements, max |
± 0.5 CN |
Diesel Fuel Pour Point |
|
Acceptable error limit when determining diesel-fuel pour points |
± 5 °C |
Kerosine content determination range in diesel fuels |
0-95 % |
Acceptable basic error limit when determining kerosene content in diesel fuels |
3% |
Motor Oil Clarity Level Measurement |
|
Motor oils clarity level measurement range |
95-100 % |
Acceptable error limit of motor oils clarity level measurement |
0.1 % |
Acceptable difference limit between motor oils clarity parallel measurements |
0.01 % |
Di-Electric Permeability Measurement |
|
POL dielectric permeability measurement range |
1–5 Unit |
Acceptable error limit of dielectric permeability measurement, max |
0.001 Unit |
Acceptable difference limit between POL dielectric permeability parallel measurements, max |
0.001 Unit |
Water in Oil Content |
|
Water-in-oil content determination range for oil products |
0-4 % |
Acceptable basic error limit when determining water-in-oil content of oil products |
0.001 % |
Measurement Time |
1–5 s |
Insufficient power supply indication operation threshold |
5.4 V |
Instrument useful life |
6 |
Overall Dimensions |
|
Electronic module |
100 x 210 x 40 mm |
Sensor #1 and #2 |
60 х 100 mm |
Weight |
680 Grams |
PORTABLE OCTANE AND CETANE ANALYZER