Detailed specification as follows:
Salient Features:
- Exterior boy is fabricated from Mild Steel material, which is powder coated in attractive shades.
- Interior chamber is also fabricated from Mirror polished Stainless Steel S.S. ‑ 304 Material
- Outer double walled metal door with sponge ‑ type silicon gasket for air‑tight sealing as to avoid leakage of chamber uniformity
- Electrical wiring as per CE Compliances ‑ with individual numbering of each internal circuitry wire for ease of maintenance & Repairing
- Chamber illumination is accomplished by Fluorescent light with door switch.
- Cord wire duly tested and inspected with stress factor as per CE standard
- Electric Motor of blower located at the back side of unit, which is protected with safety cover to avoid accident
- Standard motor of reputed companies ‑ such as CG, GODREJ or GE Make
- User friendly and tactfully designed chamber door and locking mechanism
- Aesthetic outer appearance and high quality
- Unit mounted on castor wheels for easy movement.
- Supplied with instruction Manual & Circuit Diagram
- IQ, OQ, PQ at extra cost.
Air Circulation:
Temperature is maintained by a quiet running blower for better air circulation throughout the chamber. Forced air circulated vertically down and re circulated throughout the chamber for uniform temperature distribution.
Heating & Cooling System:
Long lasting U Shaped tubular heaters used as heating element. Hermetically sealed compressor CFC free compressor (134 A gas) coupled with evaporation coil and condenser.
Trays for samples:
- Removable wire mess trays are part of the standard supply.
- Strong enough to with‑stand heavy load of test specimens, which should be equally distributed among trays.
- Such trays will also facilitate better air circulation.
- The distance between trays will be 15‑20 centimeters.
Safety Features:
- High temperature safety cut off (Thermostat)
- Electrical overload safety (Fuse 10 Amperes)
- Time delay for compressor switch ON
Pre installation mandatory requirement : (Guarantee void in case of non compliance):
- An operator shall maintain Room temperature around machine preferably at 25 °C with air conditioning system or shall arrange for a well ventilated room with exhaust fan. However, surrounding temperature should not exceed 30 °C .
- Stabilized Input Voltage of 230 Volt, 50 Hz, 20 Amps, AC Supply must be available for error less performance of the instrument. Use of Servo Controlled Stabilizer is recommended.
Construction |
Double walled |
Interior |
S.S. 304 Grade mirror polished Stainless steel material |
Exterior |
Mild steel material, which is powder coated in attractive shades |
Inner working size |
30X30X30 CM (HXWXD) |
Chamber volume |
27 Ltr |
Thermal Insulation |
PUF insulation to avoid thermal loss (B/w Inner chamber and exterior body) |
Insulation thickness |
70 mm |
Insulation source |
Imported from |
Temperature Range |
5 °C to 60 °C |
Temperature Resolution |
0.1 °C |
Temperature Accuracy |
± 0.5°C or better |
Temp Uniformity |
|
Temperature Control |
MICRO-PROCESSOR based Auto tune PID Digital Temp Indicator cum Controller |
Temperature Display |
LED Display |
Temperature sensor |
High quality PT-100 Sensor |
Illumination |
Fluorescent Tube light |
Tube Light Make |
GELCO |
Fluorescent light watts |
11 Watts |
Heating source |
U-shaped S.S. Tubular heaters |
Heater watts |
500 Watts |
Blower |
For Forced Air circulation. |
Motor of Blower |
Marathon/GE/Any other reputed make. |
No. of trays |
|
Refrigeration |
CFC free EMERSON compressor utilizing R-134a eco friendly refrigerant. |
Observation |
Inner transparent acrylic door for quick viewing and retrieval. |
Door lock |
Ergonomic and secured key lockable door lock. |
Feet On |
Castor wheels (For easy mobility) |
Electrical |
230 Volt, 15A, 50 Hz, AC Supply. |
Overall Dimensions |
|
Weight (Approx.) |