WALK IN STABILITY CHAMBER-(2000 LTR)-(1.6 X 0.75 X 1.7 METER) (W X D X H)-PLC, HMI & 21-CFR SOFTWARE

WALK IN STABILITY CHAMBER-(2000 LTR)-(1.6 X 0.75 X 1.7 METER) (W X D X H)-PLC, HMI & 21-CFR SOFTWARE

Model Number : EIE-106WGT20

None


Construction Details

  1. CE Certified
  2. Designed as per ICH guidelines to meet, WHO and USFDA requirement
  3. Fulfills storage conditions of  25° C - 60% RH, 40° C - 75% RH, 30° C - 65% RH, 25° C - 40% RH,  30° C - 35% RH
  4. Chamber equipped with PLC & Touch screen HMI system
  5. Highly accurate Individual control Loop for precise control of temp and humidity
  6. Interior fabricated from matt finished Stainless Steel  (S.S. - 304)
  7. Exterior body is fabricated from matt finished Stainless Steel (S.S. - 304)
  8. Toughened glass window to inspect samples without affecting the chamber temperature.
  9. Prefabricated Puff Panels with cam-locks to avoid thermal leakage
  10. Electrical wiring as per CE Compliances
  11. Chamber illumination by LED light with door switch.
  12. User friendly and tactfully designed chamber door and locking mechanism
  13. Facility to open chamber door from inside, in case of emergency
  14. Aesthetic outer appearance and high quality
  15. Supplied complete with IQ/OQ/PQ Documents, Calibration certificates and Instruction manual.


Following advance features are available

  1. Equipped with 7" Touch screen HMI Controller and PLC System
  2. Equipped with GSM MODULE (Mobile alert configuration) system for parameter deviation
  3. Equipped with Magnetic Door Lock Mechanism 
  4. Password protected door access system
  5. Supplied complete with 21-CFR (PART-11) Compliance PC Software
  6. Audit trail generation in PC Software
  7. Equipped with In-Built Data logger (Uniformity module) system 
  8. Also equipped with 4 Temp + 4 RH Sensor for data logging purpose
  9. Monitor Low water level status
  10. Facilitates automatic water filling system
  11. Auto resume after power failure
  12. Ready to use STANDBY REFRIGERATION SYSTEM in case main system fails
  13. Ready to use STANDBY HUMIDITY SYSTEM in case main system fails
  14. Auto Switch-Over facility from Mains to Standby system and vice versa
  15. Calendar options for choosing the days on which Either Mains or Standby system will remain active



Technical Specification

•Chamber Size: 1.6 x 0.75 x 1.7 (W x D x H) Meter (1600 X 750 X 1700 MM)
•Chamber volume: 2000 liters
•Inner chamber: Stainless steel S.S. - 304 Mirror Polished with cam lock
•Exterior body: Stainless steel S.S. - 304 Matt finished 
•Insulation: Puff insulation
•Insulation thickness: 80mm
•No. of trays: 08 Numbers
•Controlling System: Highly advanced Touch screen HMI & PLC Control system
•Display: Digital LCD Display (7" Screen Size)
•Temperature range: 10° C to 60° C
•Temperature accuracy: ± 0.5° C
•Temperature sensor: Pt - 100 sensor
•No. of Temperature + RH Sensors: 09 Nos (01 Main + 08 For data logging purpose)
•Humidity Range: 40% RH to 90% RH
•Humidity Accuracy: ± 3% RH
•Rh sensor: Capacitance type RH Sensor




•Illumination: LED Light
•Time & Date: User adjustable Date & Time settings
•PC Software: 21-CFR (Part-11) compliance software as per USFDA Guidelines
•Communication: RS - 232/RS - 485 communication port
•Magnetic door lock: Password protected entry levels for Operator, Supervisor etc
•IQ, OQ, PQ: IQ, OQ, DQ, Documents as per USFDA Guidelines, Validation Report and Instruction Manual.
•Blower: For uniform temp & humidity distribution
•Power Consumption: 5000 Watts (5 KW)
•Power Supply: 440 Volts, 50 HZ, Three Phase, AC Supply



Salient Features of 21-CFR Compliance Software

  1. Multi Chamber data - Acquisition Software
  2. Generates graphical analysis of DATA as per - 21 CFR Compliance guidelines
  3. File encryption Facility
  4. Secure login facility
  5. Password protected. System access is controlled by the system supervisor
  6. Limit system access
  7. Limit module access
  8. Password expiration
  9. Bilateral user ID and password identification
  10. Access failure lock out
  11. System access logs
  12. System audit trail report.
  13. System Input / Output  logs
  14. Mean kinetic temperature calculation
  15. Maintains log of user activities like users logged in, login/logout time, and changes made by the user in software settings Audit Trail Reports.
  16. Ensure Gap-free Records
  17. Schedule Reports via Email
  18. Meet Tough Standards
  19. View Environments in Real-Time
  20. Flexible Alarm Management
  21. Consistent Alarm Acknowledgements
  22. Highly Secure & Temper proof
  23. Protection From Data Loss 


Air Circulation

Temperature is maintained by a quiet running blower circulation throughout the chamber. Forced air circulated vertically down and re-circulated throughout the chamber for uniform temperature and humidity distribution.


Humidity Generation

Humidity created by steam injection method. The boiler tank is fitted in the back side of the chamber for better servicing. Electromagnetic switch is provided for controlling the wet heater from burning off, if water level is not adequate. Float valve is provided to automatically control & maintain water level in the boiler tank.

 

Heating & Cooling System

Long lasting SS tubular heaters used as heating element. The stainless steel fins provided to ensure better heat transfer. Hermetically sealed compressor CFC free compressor (134 A gas) coupled with evaporation coil and condenser. All refrigeration components like expansion valve, driers, oil separators, pressure gauges and vibration eliminators provided. Reputed company make refrigeration system such as Emerson/Kirloskar Make

 

Door & Gasket

Robust type hinges fitted on the right hand side of the chamber with front opening single door. Leak proof double lined silicon gasket will be provided. The inner surface of the door will be provided with pre-polished stainless steel and will be mounted such that conduction is avoided.


Trays for samples

  1. Removable Stainless steel Perforated trays
  2. Strong enough to withstand heavy load of test specimens
  3. Such trays will also facilitate better air circulation.
  4. The distance between trays will be 15-20 centimeters.
  5. Number of tray supplied : 8 Numbers

  

Electrical controls

Relays, Contactors, SSR, MCB, Mains Switch, Fan control switch, compressor ON/OFF switch with indicator, Dry heater switch with indicator for safe and ease of operation.

 

Electrical Circuitry Wires

CE Certified : Inner circuitry wires with individual number tags for ease of servicing and maintenance


Safety Provisions

  1. Over temperature thermostat & thermal fuses provided for Heating system.
  2. Low water level boiler cut off
  3. Electrical overload cut off
  4. Time delay for compressor switch ON
  5. Electrical circuit breaker


Viewing window

Multi pane window, toughened clear glasses sealed with silicone and separated by spacers to form a completely vacuumised window.

 

Wiper

Viewing window facilitates with wiper to clear the condensed water droplets on the glass.



Chamber Size

1.6 x 0.75 x 1.7 (W x D x H) Meter (1600 X 750 X 1700 MM)

Chamber volume

2000 liters

Inner chamber

Stainless steel S.S. - 304 Mirror Polished with cam lock

Exterior body

Stainless steel S.S. - 304 Matt finished

Insulation

Puff insulation

Insulation thickness

80mm

No. of trays

08 Numbers

Controlling System

Highly advanced Touch screen HMI & PLC Control system

Display

Digital LCD Display (7" Screen Size)

Temperature Range

10° C to 70° C

Temperature Accuracy

± 0.5° C

Temperature Sensors

Pt - 100 sensor

No. of Temperature + RH Sensors

09 Nos (01 Main + 08 For data logging purpose)

Humidity Range

40% RH to 90% RH

Humidity Accuracy

± 3% RH

Rh sensor

Capacitance type RH Sensor

Illumination

LED Light

Time & Date

User adjustable Date & Time settings

PC Software

21-CFR (Part-11) compliance software as per USFDA Guidelines

Communication

RS - 232/RS - 485 communication port

Magnetic door lock

Password protected entry levels for Operator, Supervisor etc

IQ, OQ, PQ

IQ, OQ, DQ, Documents as per USFDA Guidelines, Validation Report and Instruction Manual.

Blower

For uniform temp & humidity distribution

Power Consumption

5000 Watts (5 KW)

Power Supply

440 Volts, 50 HZ, Three Phase, AC Supply

INQUIRE NOW
name
company name
city
Email
Mobile number
requirments
message
Reset

Inquiry right arrow Inquiry left arrow INQUIRY

catalogue-image