- CE Certified
- Designed as per ICH guidelines to meet, WHO and USFDA requirement
- Fulfills storage conditions of 25° C - 60% RH, 40° C - 75% RH, 30° C - 65% RH, 25° C - 40% RH, 30° C - 35% RH
- Chamber equipped with PLC & Touch screen HMI system
- Highly accurate Individual control Loop for precise control of temp and humidity
- Interior fabricated from mirror polished Stainless Steel (S.S. - 304)
- Exterior body is fabricated from Stainless Steel (S.S. - 304)
- Toughened glass wondow to inspect samples without affecting the chamber temperature.
- Prefabricated Puf Panels with cam-locks to avoid thermal leakage
- Electrical wiring as per CE Compliances
- Chamber illumination by LED light with door switch.
- User friendly and tactfully designed chamber door and locking mechanism
- Facility to open chamber door from inside, in case of emergency
- Aesthetic outer appearance and high quality
- Supplied complete with IQ/OQ/PQ Documents, Calibration certificates and Instruction manual.
- Equipped with 7" Touch screen HMI Controller and PLC System
- Equipped with GSM MODULE (Mobiler alert configuration) system for parameter deviation
- Equipped with Magnetic Door Lock Mechanism
- Password protected door access system
- Supplied complete with 21-CFR (PART-11) Compliance PC Software
- Audit trail generation in PC Software
- Equipped with In-Built Data logger (Uniformity module) system
- Also equipped with 4 Temp + 4 RH Sensor for data logging purpose
- Monitor Low water level status
- Facilitates automatic water filling system
- Auto resume after power failure
- Ready to use STANDBY REFRIGERATION SYSTEM in case main system fails
- Ready to use STANDBY HUMIDITY SYSTEM in case main system fails
- Auto Switch-Over facility from Mains to Standby system and vice versa
- Calendar options for choosing the days on which Either Mains or Standby system will remain active
- Multi Chamber data - Acquisition Software
- Generates graphical analysis of DATA as per - 21 CFR Compliance guidelines
- File encryption Facility
- Secure login facility
- Password protected. System access is controlled by the system supervisor
- Limit system access
- Limit module access
- Password expiration
- Bilateral user ID and password identification
- Access failure lock out
- System access logs
- System audit trail report.
- System Input / Output logs
- Mean kinetic temperature calculation
- Maintains log of user activities like users logged in, login/logout time, and changes made by the user in software settings Audit Trail Reports.
- Ensure Gap-free Records
- Schedule Reports via Email
- Meet Tough Standards
- View Environments in Real-Time
- Flexible Alarm Management
- Consistent Alarm Acknowledgements
- Highly Secure & Temper proof
- Protection From Data Loss
- Removable Stainless steel rod trays
- Strong enough to withstand heavy load of test specimens
- Such trays will also facilitate better air circulation.
- The distance between trays will be 15-20 centimeters.
- Number of tray supplied : 66 Numbers
- Over temperature thermostat & thermal fuses provided for Heating system.
- Low water level boiler cut off
- Electrical overload cut off
- Time delay for compressor switch ON
- Electrical circuit breaker
Chamber Size |
3.70 x 3.50 x 2.50 (W x D x H) Meter |
Chamber volume |
32500 liters |
Inner chamber |
Stainless steel S.S. - 316 Mirror Polished with cam lock |
Exterior body |
Stainless steel S.S. - 304 Matt finished |
Insulation |
Puff insulation |
Insulation thickness |
75mm |
No. of trays |
66 Numbers |
Controlling System |
Highly advanced Touch screen HMI & PLC Control system |
Display |
Digital LCD Display (7" Screen Size) |
Temperature Range |
10° C to 70° C |
Temperature Accuracy |
± 0.5° C |
Temperature sensor |
Pt - 100 sensor |
No. of Temperature Sensors |
05 Nos (01 Main + 04 For data logging purpose) |
Humidity Range |
40% RH to 90% RH |
Humidity Accuracy |
± 3% RH |
Rh sensor |
Capacitance type RH Sensor |
No. of Rh Sensors |
05 Nos (01 Main + 04 For data logging purpose) |
Validation port hole |
150 mm diameter, port hole provided with removable sealed cap. |
Observation window |
To inspect test samples without disturbing chamber operation |
Illumination |
LED Light |
Time & Date |
User adjustable Date & Time settings |
PC Software |
21-CFR (Part-11) compliance software as per USFDA Guidelines |
Communication |
RS - 232/RS - 485 communication port |
Magnetic door lock |
Password protected entry levels for Operator, Supervisor etc |
IQ, OQ, PQ |
IQ, OQ, DQ, Documents as per USFDA Guidelines, Validation Report and Instruction Manual. |
Blower |
For uniform temp & humidity distribution |
Power Consumption |
3000 Watts (3 KW) |
Power Supply |
230 Volts, 50 HZ, Single Phase, AC Supply |