Compliance Standards:
ASTM D445
Salient Features:
- Cleans up to 6 tubes simultaneously
- Heated solvent vapor process
- 5 to 15 minute cleaning time
- Low solvent loss
Details on functioning of the equipment:
- The Viscometer Tube Cleaner is a fast and efficient instrument that will clean all current types of viscometer tube internally and externally.
- Up to six viscometer tubes can be suspended in the cleaning chamber.
- Solvent vapours, heated to 80°C, permeate the viscometer tubes, condensing on the internal and external surfaces, dissolving any residues from samples passed through the tube during a test.
- A water fed cooling coil at the top of the chamber causes the vapor to continually condense and return to the receiver, minimizing solvent loss.
- The use of a low temperature re-circulator (Supplied at extra cost) to chill the cooling water will improve the solvent recovery rate.
| Temperature Range | Ambient + 5 °C to 80 °C | 
| Temperature Accuracy | ± 0.5 °C | 
| Temperature Display | LED Digital display | 
| Temperature controller | Microprocessor based DIGITAL PID Controller | 
| Temperature Resolution | 0.1 °C | 
| RELEVANT STANDARD | ASTM D445 | 
| Temperature sensor | PT-100 Sensor | 
| Heater Rating | 1500 watts | 
| Weight (Approx.) | 26.500 kg | 
| Power Supply | 230 Volts, 50 Hz, Single Phase, AC Supply | 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 