CLOUD AND POUR POINT APPARATUS- (-30 C)(REFRIGERATED)-(SINGLE TANK) -(FOUR TEST)

CLOUD AND POUR POINT APPARATUS- EIE-PTLT-104R30

Model Number : EIE-PTLT-104R30

As per ASTM D97 ISO 3016,IP 15, DIN 51597, IS 1448 Part 10 & equivalent test methods. Digital temperature display with resolution of 0.1°C. Temperature control to within ± 1°C . Long life, high safety zone, stainless steel immersion heaters, Automatic low liquid level safety cut-off & drain to empty tank almost fully, Seamless copper jackets, with disks, gaskets, test jars, corks, Non CFC, future safe, environment friendly refrigeration stainless steel tanks, powder coated exterior, PUF insulation, wheels for easy mobility. Single Tank, with 4 test positions, with temperature settings with resolution of 0.1°C for tests at successively lower temperatures, with four sets of jackets, jars, disks, gaskets and corks. With minimum temperature of -30°C. Also available Model EIE-PTLT-104R80 designed for -80°C.


Compliance with following international standards:

IP 15, ASTM D97, ASTM D2500, IS 1448 Part 10, ISO 3016, NF T 60-105

Principle:

Cloud point and pour point are indicators of the lowest temperature of utility for petroleum products. The sample is periodically examined while it is being cooled in the cloud and pour point apparatus. The highest temperature at which haziness is observed (cloud point), or the lowest temperature at which movement of the oil is observed (pour point), is reported as the test result. 

Construction Details:

  1. Model : Single Tank, Floor model with castor wheels
  2. Tank designed to accommodate 4 Test jars at a time.
  3. Inner tank made of 304 grade stainless steel
  4. Outer body is fabricated of mild steel with powder coated.
  5. The top surface made stainless steel with white Teflon sheet lid Digital temperature display with resolution of 0.1° C
  6. Long life, high safety zone, stainless steel immersion heaters
  7. Stirrer will be provided for uniform cooling throughout the Bath.
  8. Supplied complete with Seamless copper jackets, with disks, gaskets, test jars and silicon rubber corks.
  9. Suitable Bath media to operate the Instrument at zero temperature has to be arranged by the user.

Other available Models

  1. EIE-PTLT-104R80(-80°C)
  2. EIE-PTLT-104RI3C(3 compartment with individual controller)
  3. EIE-PTLT-104RI4C (4 compartments with individual controller) - to suit different applications.

Insulation

PUF insulation

Temperature

With minimum temperature of ‑30°C

Temperature Accuracy

Within ± 1° C

Temperature Resolution

0.1° C

Compartment size

200 mm x 200 mm x 175 mm Height approx

Volume

6 Liters of each compartment

Compressor

Emersion make compressor with CFC Free refrigerant

Power Supply

230 V, single phase, 50 Hz

Overall height

1045 mm approx

Overall Width

600 mm approx

Overall Length

450 mm approx

Refrigeration

Hermetically sealed Emersion make compressor, overload protector, fan motor, Dan Foss made dryer, condenser which are safely placed below

Weight (Approx)

110 kg (Unpacked)

Weight (Gross)

180 kg (Packed)